SMITHS MEDICAL INTERNATIONAL, LTD. PORTEX GENERAL ANESTHESIA CIRCUITS; CIRCUIT, BREATHING (W CONNECTOR, ADAPTOR, Y PIECE)
|
Back to Search Results |
|
Catalog Number C37101330-NLJ |
Device Problem
Defective Component (2292)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 03/01/2022 |
Event Type
malfunction
|
Event Description
|
It was reported that during the pre-use check, the 15-mm connector was found deformed.No patient involvement.
|
|
Manufacturer Narrative
|
Other, other text: this remediation mdr was generated under protocol b10010116, as a result of warning letter cms#617147.No problem or issues were identified during the device history record review.Three pictures were attached and during the analysis conducted it could be observed a top view of a connecter, while the other two are made a focus on the inner diameter which is deformed.The complainant returned unit was visually inspected under normal conditions of illumination according to procedure.The inner diameter of the connector is deformed; thus failure mode is confirmed.No other analysis was required.Based on the investigation and the manufacturing process the root cause that was determined is component was received to supplier in this condition.An awareness notification was sent to supplier.
|
|