Model Number 71440192
Device Problems Material Fragmentation (1261); Naturally Worn (2988)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/01/2022
Event Type  Injury  
Manufacturer Narrative
Internal complaint reference: (b)(4).Results of investigation: the associated device was returned and evaluated.A visual inspection of the returned device confirmed the stated failure mode.The device was found to be worn along the impactor face, rendering the device inoperative.The device shows signs of extensive use.A review of complaint history revealed similar events for the listed device over the previous 12 months, but no similar events for the batch based on the historical data, this failure mode will be monitored for future complaints for any necessary corrective actions.The clinical/medical evaluation concluded that per complaint details, the consigned impactors were leaving ¿black bits¿ in a ¿very few¿ number of patients due to wear and tear.Reportedly, ¿a through washout was performed¿ and ¿no pieces of black plastic have been left inside any patients and no harm has come to any of the patients¿.The requested operative notes were reportedly unavailable.Per the cleaning, sterilization and maintenance brochure [01004_v2], end of (product) life is normally determined by wear and damage due to use, and all reusable devices should be visually inspected for damage or wear and discarded.As reported, the event was due to ¿wear and tear¿.The impact to the patients beyond that which was reported would not be anticipated as it was reported that ¿no pieces.Have been left inside any patients and no harm¿ resulted.No further medical assessment is warranted at this time.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we do not have reason to suspect that the product failed to meet any product specifications at the time of manufacture.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.Assessment of historical escalated cases concluded that there are no prior actions related to this device and failure mode.A contribution of the device to the reported event could be corroborated as the device shows signs of damage/wear.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor future complaints and investigate as necessary.We consider this investigation closed.
 
Event Description
It was reported that, during tka surgery, the genesis ii tibial baseplate impactor is leaving black bits in the patient during use, due to wear and tear.A thorough washout was performed to avoid any bits being left in the patient.Surgery was performed, without any delay, with the same device.Patient was not harmed as consequence of this problem.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GII TIBIAL BASE IMPACTOR
Type of Device
PROSTHESIS, HIP, FEMORAL COMPONENT, CEMENTED, METAL
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer (Section G)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key14246271
MDR Text Key290372372
Report Number1020279-2022-02039
Device Sequence Number1
Product Code JDG
UDI-Device Identifier03596010214386
UDI-Public03596010214386
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
K123598
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Consumer,Health Professional,User Facility,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 04/28/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/28/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date08/14/2012
Device Model Number71440192
Device Catalogue Number71440192
Device Lot Number02HM02884
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/14/2022
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/05/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Outcome(s) Other; Required Intervention;
-
-