U.S. Department of Health and Human Services

MAUDE Adverse Event Report: INTUITIVE SURGICAL, INC NONE; TIP COVER ACCESSORY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

INTUITIVE SURGICAL, INC NONE; TIP COVER ACCESSORY Back to Search Results
Model Number 400180
Device Problem Material Separation (1562)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/29/2022
Event Type  Injury  
Manufacturer Narrative
The monopolar curved scissor tip cover accessory has not been returned for evaluation; therefore, the root cause of the customer reported failure mode cannot be determined.A follow-up mdr will be submitted if the instrument is returned (post engineering evaluation) or if additional information is received.A review of the site's complaint history does not show any additional complaints related to this product and/or this event.No image or procedure video was provided for review.The lot # was not provided.Therefore, an instrument log review of the product related to the complaint cannot be performed at this time.This complaint is being reported based on the following conclusion: it was alleged that the monopolar curved scissors (mcs) tip cover accessory fell inside the patient during a da vinci assisted procedure.It is unknown if the fragment was retrieved during the same procedure with no patient injury.At this time it is unknown what caused the breakage to occur.While there was no harm or injury to the patient, the reported failure mode could likely cause or contribute to an adverse event if it were to recur as unintended fragment(s) falling into the patient may require surgical intervention.
 
Event Description
It was reported that during a da vinci-assisted benign hysterectomy surgical procedure, the monopolar curved scissors (mcs) tip cover accessory was missing.The clinical sales associate (csa) called in from off site to ask if the mcs tip cover could be picked up on x-ray.Technical support engineer (tse) informed the csa that the tip cover was radiolucent and would not be picked up by x-rays.Csa informed the tse that site was still in the middle of the procedure and planning to look for the tip cover, but wanted to know if it would show up on x-ray.Csa was unable to provide additional information due to case ongoing.Tse requested that the csa call back in if site reports that they were able to find the missing tip cover.The csa reported no other issues during the procedure.There was no report of patient injury.Intuitive surgical, inc.(isi) made multiple follow-up attempts to obtain additional information.However, no further details have been received as of the date of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NONE
Type of Device
TIP COVER ACCESSORY
Manufacturer (Section D)
INTUITIVE SURGICAL, INC
3410 central expressway
santa clara CA
Manufacturer (Section G)
INTUITIVE SURGICAL, INC
3410 central expressway
santa clara CA
Manufacturer Contact
izabel nielson
3410 central expressway
santa clara, CA 
4085232100
MDR Report Key14247686
MDR Text Key293854269
Report Number2955842-2022-11318
Device Sequence Number1
Product Code NAY
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K112263
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Company Representative
Reporter Occupation Non-Healthcare Professional
Remedial Action Other
Type of Report Initial
Report Date 03/29/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/29/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number400180
Device Catalogue Number400180
Device Lot NumberN/A
Was Device Available for Evaluation? No
Date Manufacturer Received03/29/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberN/A