Catalog Number AS18511-01
Device Problem Material Twisted/Bent (2981)
Patient Problem Rupture (2208)
Event Date 04/19/2022
Event Type  Injury  
Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Event Description
A physician reported irrigation tube was found bent and deformed it was straightened by hand and used but the anterior chamber was unstable and had a rupture during a cataract procedure.Anterior vitrectomy probe was used to complete the procedure.The same is not available for return.Additional information received from the company representative who indicated the patient's current condition is improving.
 
Manufacturer Narrative
The manufacturer internal reference number is: (b)(4).
 
Manufacturer Narrative
The customer did not retain the product lot information for this procedure pack, therefore the device history records traceable to the reported procedure pack could not be reviewed.Based on our current tracking, there are no adverse trends for this reported complaint.A customer reported that an irrigation tube of the fluidics management system (fms) pak was bent and deformed.A wet fms was visually inspected.The irrigation line was bent from the cassette insertion.The fms was tested on a console.The sample primed and tuned with the ultrasonic handpiece successfully.No message code, no fluid or air leaks, and no cracks on the luers were found.The bend on the irrigation line did not affect the functionality.While the customer¿s complaint of bent tubing was confirmed, testing of the sample showed that the bend was cosmetic and did not affect the performance of the product.The cause of the bend is consistent with a defect observed when the tubing is improperly placed in the tray during the manufacturing process.Procedural enhancements will be implemented for better handling of tubing and will help mitigate and reduce recurrence of the reported issue.All lots are verified that all required inspections have been performed and all acceptance criteria are met prior to release.Quality assurance has reviewed this complaint and will continue to monitor data for evidence of adverse trending and take further action, as appropriate.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INFINITI SURGICAL PROCEDURE PAK
Type of Device
GENERAL SURGERY TRAY (KIT)
Manufacturer (Section D)
ALCON RESEARCH, LLC - HOUSTON
9965 buffalo speedway
houston TX 77054
Manufacturer (Section G)
ALCON RESEARCH, LLC - HOUSTON
9965 buffalo speedway
houston TX 77054
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key14250634
MDR Text Key290383890
Report Number1644019-2022-00326
Device Sequence Number1
Product Code LRO
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K880961
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 07/28/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/29/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue NumberAS18511-01
Device Lot NumberASKU
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Date Manufacturer Received07/21/2022
Was Device Evaluated by Manufacturer? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-