• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SIRONA DENTAL SYSTEMS GMBH MIDWEST STYLUS ATC HANDPIECE, AIR-POWERED, DENTAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SIRONA DENTAL SYSTEMS GMBH MIDWEST STYLUS ATC HANDPIECE, AIR-POWERED, DENTAL Back to Search Results
Catalog Number 882300
Device Problem Detachment of Device or Device Component (2907)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Event Description
This report summarizes 1 malfunction event where a midwest stylus atc handpiece would not hold burs. No injury resulted in the event.
 
Manufacturer Narrative
1 of 1 device was returned for evaluation. Evaluation of 1 device found chuck wear and lack of maintenance. The device was cleaned of debris and the turbine was replaced. The device was repaired and returned to the customer. This event is being submitted as part of vmsr. Only one event was received for this device during this quarter.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameMIDWEST STYLUS ATC
Type of DeviceHANDPIECE, AIR-POWERED, DENTAL
Manufacturer (Section D)
SIRONA DENTAL SYSTEMS GMBH
fabrikstrasse 31
bensheim, hessen D-646 25
GM D-64625
Manufacturer (Section G)
SIRONA DENTAL SYSTEMS GMBH
fabrikstrasse 31
bensheim, hessen D-646 25
GM D-64625
Manufacturer Contact
hannah seevaratnam
221 w. philadelphia st.
suite 60w
york, PA 17401
7178494593
MDR Report Key14251676
MDR Text Key290453614
Report Number9614977-2022-00030
Device Sequence Number1
Product Code EFB
Combination Product (y/n)N
PMA/PMN Number
K003518
Number of Events Reported1
Summary Report (Y/N)Y
Report Source Manufacturer
Reporter Occupation
Type of Report Initial
Report Date 04/29/2022
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received04/29/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue Number882300
Was Device Available for Evaluation? No Answer Provided
Was Device Evaluated by Manufacturer? No Answer Provided

-
-