Model Number SN6CWS
Device Problem Scratched Material (3020)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/27/2022
Event Type  malfunction  
Manufacturer Narrative
The product was not returned for evaluation.All product and batch history records are quality reviewed prior to product release.Information was provided that indicated the use of a non-qualified viscoelastic.The root cause for the reported complaint may be related to a failure to follow the instruction for use (ifu).A non-qualified viscoelastic was used.Material properties of a non-qualified viscoelastic may contribute to underfill, overfill, misfolding of the trailing haptic, lack of lubricity or other unpredictable outcomes.There have been one other complaints reported in the lot number.The manufacturer internal reference number is: (b)(4).
 
Event Description
A physician reported during an intraocular lens implant procedure, there was crease on optic found in eye when viewed under microscope.The surgeon removed the lens from the eye without any patient harm during the initial procedure.Additional information has been requested.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACRYSOF IQ SINGLEPIECE IOL W/ACRYSERT C DELIVERY SYSTEM
Type of Device
LENS, GUIDE, INTRAOCULAR
Manufacturer (Section D)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer (Section G)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key14285343
MDR Text Key290780386
Report Number1119421-2022-00930
Device Sequence Number1
Product Code KYB
UDI-Device Identifier00380652249287
UDI-Public00380652249287
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
P930014
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 05/04/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/05/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberSN6CWS
Device Catalogue NumberSN6CWS.235
Device Lot Number12751370
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/06/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/24/2019
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-