Model Number 21-7302-24
Device Problems Fluid/Blood Leak (1250); Leak/Splash (1354)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/01/2021
Event Type  malfunction  
Manufacturer Narrative
No product sample was received (photo of the product was provided); therefore, visual and functional testing could not be performed.The root cause of the reported issue was found to be that the bag sealer machine was causing a weak seal condition on the joint of the tube with the bag.A secondary root cause is that procedural testing was not followed per requirement.A device history record (dhr) review was conducted which indicated all inspections were completed and no issues were noted during manufacture.
 
Event Description
It was reported that the cassettes are leaking during filling.No patient was involved.
 
Manufacturer Narrative
Other, other text: while performing a review of submitted mdr report it was discovered that this file was inadvertently assessed as reportable.No patient death, serious injury, and no evidence of a reportable product malfunction.The hazardous situation for the given complaint device would not likely cause or contribute to a death or serious injury if the malfunction were to recur.Report 3012307300-2022-07783 is no longer considered reportable, please disregard any mdr reports associated with it.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CADD MEDICATION CASSETTE RESERVOIRS WITH FLOW STOP
Type of Device
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
ST PAUL
1265 grey fox rd.
st. paul MN 55112
Manufacturer (Section G)
NULL
1265 grey fox rd.
st. paul MN 55112
Manufacturer Contact
jim vegel
6000 nathan lane north
minneapolis, MN 55442
MDR Report Key14308551
MDR Text Key290938311
Report Number3012307300-2022-07783
Device Sequence Number1
Product Code FPA
UDI-Device Identifier10610586027239
UDI-Public10610586027239
Combination Product (y/n)N
Reporter Country CodeNL
PMA/PMN Number
K040636
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Distributor
Reporter Occupation Other
Remedial Action Recall
Type of Report Initial,Followup
Report Date 06/06/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/06/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number21-7302-24
Device Catalogue Number21-7302-24
Device Lot Number3915963
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received06/03/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/20/2020
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberZ-0148-2022
Patient Sequence Number1
-
-