• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE CORPORATION PRISMAFLEX SETS (M); DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE CORPORATION PRISMAFLEX SETS (M); DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM Back to Search Results
Catalog Number 955467
Device Problem Fluid/Blood Leak (1250)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/14/2022
Event Type  malfunction  
Event Description
It was reported that during treatment with a prismaflex m150 set, an external fluid leak was observed from the effluent line.There was no report of patient injury or medical intervention associated with this event.No additional information is available.
 
Manufacturer Narrative
The actual device was not available; however, photographs of the sample were provided for evaluation.Visual inspection of the provided pictures visual inspection did not identify any abnormalities that could have contributed to the reported condition.The reported condition was not verified.The cause of the condition could not be determined.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRISMAFLEX SETS (M)
Type of Device
DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
BAXTER HEALTHCARE - MEYZIEU
7, av lionel terray, b.p. 126
meyzieu cedex rhone 69883
FR   69883
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 
2242702068
MDR Report Key14317740
MDR Text Key291393086
Report Number8010182-2022-00117
Device Sequence Number1
Product Code KDI
UDI-Device Identifier07332414123451
UDI-Public(01)07332414123451
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Nurse
Type of Report Initial
Report Date 05/06/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/07/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/31/2023
Device Catalogue Number955467
Device Lot Number21F3003
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/18/2022
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-