Product event summary: the 4fc12 sheath with lot 0010663031 was returned and analyzed.
Visual inspection prior to disassembly and functional testing was performed on the shaft, handle, and dilator.
No anomalies were identified.
The handle has no cosmetic issue and side tube is connected properly to the handle.
The shaft had multiple kinks approximately 2 inches from tip.
The steering mechanism of the sheath was tested and no anomalies were noted in the handle segment, it is plausible that due to the possible kink the user might have reported the steering issue as the shaft would not deflect properly.
In conclusion, the sheath failed the returned product inspection due to a shaft kink.
If information is provided in the future, a supplemental report will be issued.
|