Catalog Number 08D06-77
Device Problem False Negative Result (1225)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/28/2022
Event Type  malfunction  
Manufacturer Narrative
An evaluation is in process. A follow-up report will be submitted when the evaluation is complete.
 
Event Description
The customer observed false nonreactive architect syphilis tp results on one 59yrs old female patient. The results provided were: on 28apr2022
=
0. 39 s/co (< 1. 00 s/co
=
nonreactive) /patient history
=
4. 67 s/co (positive) /tppa
=
positive there was no reported impact to patient management.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameARCHITECT SYPHILIS TP REAGENT KIT
Type of DeviceENZYME LINKED IMMUNOABSORPTION ASSAY, TREPONEMA PALLIDUM
Manufacturer (Section D)
ABBOTT GMBH
max-planck-ring 2
wiesbaden 65205
GM 65205
Manufacturer (Section G)
ABBOTT GMBH
max-planck-ring 2
wiesbaden 65205
GM 65205
Manufacturer Contact
siobhan wright
lisnamuck
post market surveillance
longford N39 E-932
EI   N39 E932
433331157
MDR Report Key14321175
MDR Text Key291721767
Report Number3002809144-2022-00153
Device Sequence Number1
Product Code LIP
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K153730
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 05/19/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/07/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Expiration Date06/21/2022
Device Catalogue Number08D06-77
Device Lot Number30344BE01
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received05/19/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/15/2021
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unkown

Patient Treatment Data
Date Received: 05/07/2022 Patient Sequence Number: 1
Treatment
ARC I2000SR INST, 03M74-02, ISR64239.
-
-