• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY MITEK LLC US MICRO TORNADO HP W HANDCONTROL; LINE-POWERED ARTHROSCOPIC SHAVER SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY MITEK LLC US MICRO TORNADO HP W HANDCONTROL; LINE-POWERED ARTHROSCOPIC SHAVER SYSTEM Back to Search Results
Model Number 283512
Device Problems Electrical /Electronic Property Problem (1198); Use of Device Problem (1670); Electrical Shorting (2926)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/15/2022
Event Type  malfunction  
Event Description
It was reported by the affiliate in (b)(6) that the micro tornado hp w handcontrol device did not work anymore.During in-house engineering evaluation, it was determined that the there was short circuit in the motor cable on the device.There was no procedure nor patient involvement reported.No additional information was provided.
 
Manufacturer Narrative
If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Udi: (b)(4).The complaint device was received at the service center and evaluated.It was reported that device do not work anymore.Per service reports, this complaint can be confirmed.During the service evaluation the following defects were identified: the buttons have no function and the motor is stuck.There is a short circuit in the cable.The defective hand control set , defective motor cable were replaced to resolve the issues.After repair, the device was found to be working according to the specifications.The faulty parts was identified as the root cause for the device failure during the service evaluation.Manufacturing record evaluation is not required as the reported event is not associated with the manufacturing process and/or the potential cause of the defect cannot be associated to manufacturing.At this point in time, no corrective action is required, and no further action is warranted.Depuy mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.
 
Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by mitek or its employees that the report constitutes an admission that the device, mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H4: the device manufacture date was unknown in the initial report and has been updated accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MICRO TORNADO HP W HANDCONTROL
Type of Device
LINE-POWERED ARTHROSCOPIC SHAVER SYSTEM
Manufacturer (Section D)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
Manufacturer (Section G)
JTE WARSAW MFG SITE
700 orthopaedic drive
warsaw IN 46582
Manufacturer Contact
kate karberg
325 paramount drive
raynham, MA 02767
3035526892
MDR Report Key14347430
MDR Text Key291517944
Report Number1221934-2022-01377
Device Sequence Number1
Product Code HRX
UDI-Device Identifier10886705016938
UDI-Public10886705016938
Combination Product (y/n)N
Reporter Country CodeSZ
PMA/PMN Number
K954465
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 05/10/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/10/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number283512
Device Catalogue Number283512
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/17/2022
Is the Reporter a Health Professional? No
Date Manufacturer Received06/10/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/21/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Removal/Correction NumberN/A
Patient Sequence Number1
-
-