Model Number 1012-04-050 |
Device Problem
Adverse Event Without Identified Device or Use Problem (2993)
|
Patient Problems
Foreign Body Reaction (1868); Pain (1994); Loss of Range of Motion (2032); Discomfort (2330); Osteolysis (2377); Ambulation Difficulties (2544); Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581)
|
Event Date 06/28/2021 |
Event Type
Injury
|
Event Description
|
Pinnacle claim submission form and medical records received.After review of the medical records the patient was revised to address osteolysis and metallosis resulting to hip pain and walking difficulty.Operative note reported there was a fair amount of grayish brownish synovial fluid consistent with metallosis, cup is slight vertical position.Clinical visit reported right hip pain, discomfort, limited adl, antalgic gait and elevated metal ions.Doi: (b)(6) 2009; dor: (b)(6) 2021; right hip.
|
|
Manufacturer Narrative
|
Product complaint # (b)(4).Initial reporter occupation: lawyer.(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Manufacturer Narrative
|
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
|
Manufacturer Narrative
|
Product complaint # (b)(4).Investigation summary = > no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot = > a manufacturing record evaluation (mre), was not possible because the required lot code was not provided.
|
|
Search Alerts/Recalls
|