Model Number HST III SYSTEM (3.8MM) |
Device Problems
Fitting Problem (2183); Activation Problem (4042)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 03/16/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
Trackwise id (b)(4).Since the device is not available to be returned to us, a technical evaluation cannot be performed.Per our standard sop's, all events are tracked and trended to determine whether or not any trends develop.
|
|
Event Description
|
Related to (b)(4).The hospital reported that the hst iii system (3.8mm) occluder does not release.The diameter of the blocking umbrella is larger than the diameter of the delivery tube, so it cannot be released.The blocking umbrella is in the delivery pipe.No surgical delay.Replace the same equipment to complete the surgery.No harm to the patient.
|
|