A user facility biomedical technician (biomed) reported that an aquabplus 2000 sustained thermal damage.
There were no alarms.
The aquabplus system was in standby mode at the time of the discovery.
The biomed stated they were replacing the level rod (with float) because the disinfection cycle would not complete.
Upon lifting the cover, a burnt wire connector was identified on the stage 1 power switch.
This occurred after the clinic had closed for the day and there was no patient involvement.
There was no evidence of any burning smell, smoke, sparks, or flames.
The biomed stated there was nothing wrong with the machine, other than the fact that it would not complete the disinfection cycle.
The thermal overload switch was not tripping, and there were no blown fuses in the local power supply.
The biomed also confirmed there were no local power grid issues at the time of the event.
After replacing the level rod, the machine was able to complete disinfection.
The burnt wire connector was not replaced.
The biomed was unable to provide the ftp files.
The level rod that was replaced was available to be returned for evaluation.
|