• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SIEMENS AG/SIEMENS HEALTHCARE GMBH SIEMENS ESPREE; SYSTEM, NUCLEAR MAGNETIC RESONANCE IMAGING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SIEMENS AG/SIEMENS HEALTHCARE GMBH SIEMENS ESPREE; SYSTEM, NUCLEAR MAGNETIC RESONANCE IMAGING Back to Search Results
Device Problem Improper or Incorrect Procedure or Method (2017)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/28/2022
Event Description
A patient entered the mri scan room with a crutch that was pulled out of the patient's hand and flew against the magnet with sufficient force to crack the crutch.No patients or staff were injured but the facility mri screening and safety procedures were not followed and the facility was found not to have a ferromagnetic detection screening device to test objects for ferrous properties.Fda safety report id# (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SIEMENS ESPREE
Type of Device
SYSTEM, NUCLEAR MAGNETIC RESONANCE IMAGING
Manufacturer (Section D)
SIEMENS AG/SIEMENS HEALTHCARE GMBH
MDR Report Key14408416
MDR Text Key291837212
Report NumberMW5109709
Device Sequence Number1
Product Code LNH
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Voluntary
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 05/11/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/13/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Device Operator Health Professional
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Other;
Patient SexFemale
-
-