• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. PUMP MMT-1714K 630G BLACK MMOL CANADA; ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. PUMP MMT-1714K 630G BLACK MMOL CANADA; ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND Back to Search Results
Model Number MMT-1714K
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Hyperglycemia (1905)
Event Date 05/10/2022
Event Type  Injury  
Event Description
It was reported that customer experienced high blood glucose level.Customer's blood glucose value was 28 mmol/l at the time of the incident.The customer was assisted with troubleshooting for high blood glucose.The customer was treated with pump.Customer had been using insulin pump system within 48 hours of reported high blood glucose event.No further complications were reported.The customer will continue to use the product.
 
Manufacturer Narrative
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.No conclusion can be drawn at this time.We therefore consider this report complete to the best of our knowledge.
 
Manufacturer Narrative
A complete analysis and testing of the insulin pump showed that, p-cap locked in place properly during testing.Unit was successfully downloaded using (thump software).The adapt tool was utilized to search for alarms that may have occurred in the past or during complain call that might of trigger the reason complain.Unit passed the displacement test, rewind test, prime/seating test, basic occlusion test, force sensor test, occlusion test and self test.During download review no relevant alarms confirm in download history files to confirm complain code.Proceeded by cutting unit open and perform a visual inspection of connectors and electronic assemblies.Per visual inspection no moisture damage or anomalies noted during visual inspection.No physical damage noted during visual inspection.In conclusion, customer concerns are not confirmed.Unable to confirm customer alleged concern of high bgs.No relevant alarms noted in download history review and testing of the unit was successful.Medtronic, inc.(medtronic) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information in the time allotted and has provided as much information as is available to the company as of the submission date this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employees caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any "defects" or has "malfunctioned".These words are included in the fda 3500a form and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.Medtronic objects to the use of these words and others like it because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PUMP MMT-1714K 630G BLACK MMOL CANADA
Type of Device
ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
Manufacturer Contact
tricha miles
ceiba norte ind. park #50 road
juncos 00777--386
7635140379
MDR Report Key14423912
MDR Text Key291938476
Report Number2032227-2022-191066
Device Sequence Number1
Product Code OZO
UDI-Device Identifier000000763000315702
UDI-Public(01)000000763000315702
Combination Product (y/n)N
Reporter Country CodeCA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 03/14/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/17/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Model NumberMMT-1714K
Device Catalogue NumberMMT-1714K
Device Lot NumberHG60429
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Date Manufacturer Received02/22/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/14/2022
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Treatment
FRN-UNK-RSVR, UNOMED INF SET
Patient Outcome(s) Other;
-
-