CONCORD MANUFACTURING 2008T HD SYS. CDX W/BIBAG BLUE STAR DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM Back to Search Results
Model Number 191126
Device Problem No Audible Alarm (1019)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/04/2022
Event Type  malfunction  
Event Description
A materials manager reported that a dialyzer blood leak occurred during a hemodialysis (hd) treatment. In a follow up with a clinic manager, it was clarified that the blood leak occurred at one hour and three minutes into the hd treatment. The blood leak was described as being an internal blood leak that was visually seen at the hansen¿s connector. Blood tests strips were used and tested positive for the presence of blood. The machine was a fresenius 2008t machine. The machine did not alarm. The machine was removed from the floor and the treatment was completed with another machine and new supplies. There was no damage noted on the dialyzer prior to use. The patient¿s estimated blood loss (ebl) was greater than 100 ml. There was no patient injury, adverse events, or medical intervention required as a result of this event. The treatment was interrupted but the patient was able to complete treatment without any further events. The machine is currently still being evaluated.
 
Manufacturer Narrative
The plant investigation is in process. A supplemental mdr will be submitted upon completion of this activity.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name2008T HD SYS. CDX W/BIBAG BLUE STAR
Type of DeviceDIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM
Manufacturer (Section D)
CONCORD MANUFACTURING
director, quality systems
4040 nelson avenue
concord CA 94520
Manufacturer (Section G)
CONCORD MANUFACTURING
director, quality systems
4040 nelson avenue
concord CA 94520
Manufacturer Contact
jason busch
920 winter st
waltham, MA 02451
9043166958
MDR Report Key14434268
MDR Text Key293644752
Report Number0002937457-2022-00793
Device Sequence Number1
Product Code KDI
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K173972
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 06/15/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/19/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model Number191126
Device Catalogue Number191126
Was Device Available for Evaluation? No
Device AgeMO
Date Manufacturer Received06/06/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured04/26/2021
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

Patient Treatment Data
Date Received: 05/19/2022 Patient Sequence Number: 1
Treatment
FRESENIUS BLOODLINES.; OPTIFLUX NRE DIALYZER.
-
-