MAUDE Adverse Event Report: NULL PORTEX GENERAL ANESTHESIA CIRCUITS; CIRCUIT, BREATHING (W CONNECTOR, ADAPTOR, Y PIECE)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NULL PORTEX GENERAL ANESTHESIA CIRCUITS; CIRCUIT, BREATHING (W CONNECTOR, ADAPTOR, Y PIECE) Back to Search Results
Catalog Number CA24B0/400/000JP
Device Problem Material Puncture/Hole (1504)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/01/2022
Event Type  malfunction  
Event Description
It was reported that during the pre-use check, a tear was found in the breathing circuit.No additional information is available.
 
Manufacturer Narrative
Other, other text: h6: event problem and evaluation codes: updated.H10: device evaluation: the device was returned for investigation.A visual inspection was performed.Five (5) pictures were attached, during the analysis conducted it could be observed a top view of the circuit product.The other pictures make focus to a section marked from the sample where it can be observed damage in circuit.Visual inspection: sample presented a tier damage in the circuit; the failure mode reported is confirmed.Per trend review an escalation was performed to investigate this failure and determine the root cause through a capa (corrective and preventive action).The cause of the reported problem was traced to the manufacturing process.A dhr (device history review) was performed and no problems or issues were identified during this dhr review.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PORTEX GENERAL ANESTHESIA CIRCUITS
Type of Device
CIRCUIT, BREATHING (W CONNECTOR, ADAPTOR, Y PIECE)
Manufacturer (Section G)
NULL
MDR Report Key14435315
MDR Text Key292028125
Report Number3012307300-2022-09090
Device Sequence Number1
Product Code CAI
Combination Product (y/n)N
Reporter Country CodeJA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 11/17/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/19/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberCA24B0/400/000JP
Device Lot Number220316
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received11/11/2022
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-