NULL PORTEX GENERAL ANESTHESIA CIRCUITS; CIRCUIT, BREATHING (W CONNECTOR, ADAPTOR, Y PIECE)
|
Back to Search Results |
|
Catalog Number CA24B0/400/000JP |
Device Problem
Material Puncture/Hole (1504)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 04/01/2022 |
Event Type
malfunction
|
Event Description
|
It was reported that during the pre-use check, a tear was found in the breathing circuit.No additional information is available.
|
|
Manufacturer Narrative
|
Other, other text: h6: event problem and evaluation codes: updated.H10: device evaluation: the device was returned for investigation.A visual inspection was performed.Five (5) pictures were attached, during the analysis conducted it could be observed a top view of the circuit product.The other pictures make focus to a section marked from the sample where it can be observed damage in circuit.Visual inspection: sample presented a tier damage in the circuit; the failure mode reported is confirmed.Per trend review an escalation was performed to investigate this failure and determine the root cause through a capa (corrective and preventive action).The cause of the reported problem was traced to the manufacturing process.A dhr (device history review) was performed and no problems or issues were identified during this dhr review.
|
|