MEDTRONIC PUERTO RICO OPERATIONS CO. SYNCHROMED II PUMP, INFUSION, IMPLANTED, PROGRAMMABLE
|
Back to Search Results |
|
Model Number 8637-40 |
Device Problems
Migration or Expulsion of Device (1395); Infusion or Flow Problem (2964)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 05/13/2022 |
Event Type
Injury
|
Event Description
|
Information was received from multiple sources (manufacturer representative, healthcare provider) regarding a patient who was receiving compounded baclofen (500 mcg at unknown dose)) via an implantable pump for intractable spasticity.
It was reported that after failed aspirating attempt, they noticed that the catheter was sutured to the pump.
Once the catheter was freed from suture, aspiration was attempted again and failed.
They then went to the back where they noticed that the catheter was not in the intrathecal space.
The catheter had migrated out of the space.
At that time ((b)(6) 2022), they revised the catheter.
There were no known factors that may have led or contributed to the issue.
The issue was resolved at the time of report.
The patient's weight and medical history were asked but unknown.
The patient's status at the time of report was alive - no injury.
|
|
Manufacturer Narrative
|
Concomitant medical products: product id: 8780, serial# (b)(4), implanted: (b)(6) 2021 explanted: (b)(6) 2022, product type: catheter.
Other relevant device(s) are: ubd: 29-oct-2022, udi#: (b)(4).
If information is provided in the future, a supplemental report will be issued.
|
|
Search Alerts/Recalls
|
|
|