H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The date of event was not provided by the complainant/reporter, the date reflected in this report is the date bd became aware of the event.The following were reviewed as part of this investigation: patient severity, applicable previous investigation(s), sample (if available), applicable fmea documents, labeling, and applicable manufacture records.Based on a review of this information, the following was concluded: the site rite prevue was visually inspected upon receipt and was found to be in poor physical condition.The reported issue that the unit will occasionally turn off by itself is unconfirmed.The unit ran for 3+ hours on battery power until the battery was depleted, which is a normal runtime.
|