Catalog Number 532.101
Device Problems Noise, Audible (3273); Physical Resistance/Sticking (4012)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/01/2022
Event Type  malfunction  
Manufacturer Narrative
This report is being submitted pursuant to the provisions of 21 cfr, part 803. This report may be based on information which has not been investigated or verified prior to the required reporting date. This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report. Device was used for treatment, not diagnosis. If information is obtained that was not available for this medwatch, a supplemental medwatch report will be filed as appropriate. The actual device was returned for evaluation. During repair, an evaluation was performed and it was determined that the device was making a strange noise and had a sticky tricky. Therefore, the reported condition was confirmed. The assignable root cause of this condition was determined to be traced to component failure due to wear. Udi ¿ (b)(4).
 
Event Description
It was reported from japan that during service and evaluation, it was determined that the upper and lower triggers of the small battery drive device hooked/sticking, the device was making a strange noise, had vibration, and had insufficient/low power. It was further determined that the device failed pretest for check for sticky triggers and check the power with the power test bench. It was noted in the service order that the device was making an abnormal noise. This event did not occur during surgery. There was no patient involvement. There were no reports of injuries, medical intervention, or prolonged hospitalization. The exact date of the event was not reported, however, it was reported that the event occurred in 2022. All available information has been disclosed. If additional information should become available, a supplemental medwatch will be submitted accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameCOLIBRI II HANDPIECE
Type of DeviceINSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT
Manufacturer (Section D)
SYNTHES GMBH
eimattstrasse 3
oberdorf 4436
SZ 4436
Manufacturer Contact
kate karberg
eimattstrasse 3
oberdorf 4436
SZ   4436
3035526892
MDR Report Key14453610
MDR Text Key300330522
Report Number8030965-2022-03318
Device Sequence Number1
Product Code HWE
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility,Company Representative
Reporter Occupation
Type of Report Initial
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received05/20/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Catalogue Number532.101
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/06/2022
Date Manufacturer Received05/11/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/29/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unkown

-
-