Model Number N/A
Device Problems Material Deformation (2976); Difficult or Delayed Separation (4044)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/27/2022
Event Type  malfunction  
Manufacturer Narrative
The information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The manufacturer has received the sample and will evaluate.Results are expected soon.A lot history review (lhr) of refu3904 showed no other similar product complaint(s) from this lot number.
 
Event Description
It was reported that during catheter placement, the customer found evidence of split at the black tip of the introducer sheath when preparing to insert the introducer sheath.
 
Manufacturer Narrative
H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The following were reviewed as part of this investigation: previous complaints, patient severity, applicable manufacturing records, submitted photograph, and applicable fmea documents.Based on a review of this information, the following was concluded: the complaint of damage to an introducer sheath was confirmed but the exact cause was unknown.A photograph was provided which depicted ptfe sheath tubing.The distal edge of the sheath contained some waviness and the sheath had been partially split at the distal tip.No observable evidence suggested the cause of this event.A review of the manufacture records associated with the provided lot number found no potential cause identified during product manufacture.Should the physical sample be returned at a later time, this investigation will be updated.
 
Event Description
It was reported that during catheter placement, the customer found evidence of split at the black tip of the introducer sheath when preparing to insert the introducer sheath.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNIVERSAL MICROINTRODUCER KIT PTFE PEEL-APART SHEATH INTRODUCER 4.5F X 5CM
Type of Device
INTRODUCER, CATHETER
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer (Section G)
BARD REYNOSA S.A. DE C.V. -9617592
blvd. montebello #1
parque industrial colonial
reynosa, tamaulipas 88780
MX   88780
Manufacturer Contact
johanna de oliveira
605 north 5600 west
salt lake city 84116
8015950700
MDR Report Key14461925
MDR Text Key292355849
Report Number3006260740-2022-01863
Device Sequence Number1
Product Code DYB
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K915167
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 06/25/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/20/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number0668945
Device Lot NumberREFU3904
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/28/2022
Is the Reporter a Health Professional? Yes
Date Manufacturer Received06/24/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/01/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-