The customer reported to olympus, during a bronchoscopy, the subject device fractured in the patient's airway when sampling a station 11r lymph node.The device fragment was 2cm long and successfully retrieved from the patient using olympus endojaw alligator model fb-211d forceps.There was a two (2) minute procedural delay when the device broke and retrieved while the patient was under moderate sedation.The intended procedure was cancelled.A post-operative chest x-ray confirmed no retained fragments and no pneumothorax.The patient was asymptomatic post-operatively.The subject device was not inspected prior to use.
|
This supplemental report is being submitted to provide the results of the legal manufacturer¿s investigation and the results of the device history records (dhr) review.The dhr for this device were reviewed and all records indicated the product was manufactured according to all applicable procedures and met final product release criteria.No abnormalities were found.A definitive root cause was not identified.Based on the available information, the legal manufacturer determined the probable cause of the failure is likely due to excessive force based on previous complaints.Content of the instruction manual (drawing number: rk1001, revision number :4) was confirmed.The instruction manual contains the following descriptions, and it warns against this event.·when inserting the instrument into the endoscope, the distal end of the needle tube may be bent.When piercing the target, confirm the distal end of the sheath and needle tube in the endoscopic field of view and/or ultrasound image while considering such bending.Otherwise, patient injury such as perforation, bleeding, or mucous membrane damage may occur.Do not try to straighten a bent or deformed needle with your hands because the needle may break.Use a spare needle instead.Olympus will continue to monitor field performance for this device.
|