• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: GLOBUS MEDICAL, INC. CREO MIS; CREO MIS LOCKING CAP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

GLOBUS MEDICAL, INC. CREO MIS; CREO MIS LOCKING CAP Back to Search Results
Model Number 1134.0010
Device Problem Device Slipped (1584)
Patient Problem Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581)
Event Date 04/15/2022
Event Type  malfunction  
Event Description
It was reported there was a revision surgery to replace (4) creo mis locking caps that were loosened post-operatively.This event occurred in japan.
 
Manufacturer Narrative
The devices were returned for evaluation.Initial observation shows all four locking caps have a small amount of surface wear.There is no excessive damage/deformation to the threads or the mating surfaces.An exact cause to the reported issue could not be determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CREO MIS
Type of Device
CREO MIS LOCKING CAP
Manufacturer (Section D)
GLOBUS MEDICAL, INC.
2560 general armistead avenue
audubon PA 19403
Manufacturer (Section G)
GLOBUS MEDICAL, INC.
2560 general armistead avenue
audubon PA 19403
Manufacturer Contact
daniel paul
2560 general armistead avenue
audubon, PA 19403
6109301800
MDR Report Key14462956
MDR Text Key300349820
Report Number3004142400-2022-00072
Device Sequence Number1
Product Code MNH
UDI-Device Identifier00889095071139
UDI-Public00889095071139
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K124058
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 05/12/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/20/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number1134.0010
Device Lot NumberBAX537FB
Was Device Available for Evaluation? Yes
Date Manufacturer Received04/25/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/15/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age29 YR
Patient SexMale
Patient EthnicityNon Hispanic
Patient RaceAsian
-
-