SMITH & NEPHEW, INC. DYONICS 25 INFLOW ONLY TUBE SET (3); ARTHROSCOPE Back to Search Results
Model Number 7211004
Device Problems Increase in Pressure (1491); Connection Problem (2900)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/01/1901
Event Type  malfunction  
Manufacturer Narrative
Internal complaint reference: case (b)(4).
 
Event Description
It was reported that, during a acl surgery, three (3) dyonics 25 cartridge tubing sets were not sitting properly in the pump so the pump is over pumping fluid.The cartridge could be manually pushed in further to solve the issue but came back out as soon as it is not being pushed.Surgery was resumed, after a non-significant delay, with a back-up device.No patient complications were reported.
 
Manufacturer Narrative
The reported device was not returned to the designated complaint unit for independent evaluation, thus visual inspection and functional testing could not be performed.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specifications upon release for distribution.A complaint history review found similar reported events.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.There was no relationship found between the device and the reported event.The complaint was not confirmed.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.No containment or corrective actions are recommended at this time.If the product associated with this event is returned at a future date, this investigation will be reopened for evaluation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DYONICS 25 INFLOW ONLY TUBE SET (3)
Type of Device
ARTHROSCOPE
Manufacturer (Section D)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107
Manufacturer (Section G)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key14478608
MDR Text Key292747397
Report Number1643264-2022-00174
Device Sequence Number1
Product Code HRX
UDI-Device Identifier03596010543561
UDI-Public03596010543561
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K051326
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 12/07/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/23/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number7211004
Device Catalogue Number7211004
Device Lot Number50976084
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/06/2022
Date Device Manufactured11/02/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-