Model Number L331
Device Problem Difficult to Insert (1316)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/04/2022
Event Type  Injury  
Event Description
It was reported that this implantable pacemaker experienced difficulty in the insertion of its leads.Multiple attempts were performed during the initial procedure.Eventually the procedure was completed and x-rays were performed for evaluation.This confirmed the issue was still present.An additional procedure was performed in order to resolve the event.This device remains in service.No further adverse patient effects were reported.
 
Manufacturer Narrative
This product has not been returned to boston scientific, and as a result, laboratory analysis could not be conducted.The associated investigation noted evidence indicating difficulty may have been encountered fully inserting an is-1 lead into the header of this device resulting in the additional procedure.Please refer to the complaint description for more information regarding the specific circumstances of this event.
 
Event Description
It was reported that this implantable pacemaker experienced difficulty in the insertion of its leads.Multiple attempts were performed during the initial procedure.Eventually the procedure was completed and x-rays were performed for evaluation.This confirmed the issue was still present.An additional procedure was performed in order to resolve the event.This device remains in service.No further adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACCOLADE MRI EL DR
Type of Device
IMPLANTABLE PACEMAKER
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key14493669
MDR Text Key292567549
Report Number2124215-2022-14079
Device Sequence Number1
Product Code LWP
UDI-Device Identifier00802526559266
UDI-Public00802526559266
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P150012/S000
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 05/26/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/25/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date01/12/2024
Device Model NumberL331
Device Catalogue NumberL331
Device Lot Number973680
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received05/23/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/03/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age60 YR
Patient SexFemale
-
-