• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. SYNCHROMED II; PUMP, INFUSION, IMPLANTED, PROGRAMMABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. SYNCHROMED II; PUMP, INFUSION, IMPLANTED, PROGRAMMABLE Back to Search Results
Model Number 8637-20
Device Problems Misconnection (1399); Obstruction of Flow (2423)
Patient Problem Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581)
Event Date 05/21/2022
Event Type  Injury  
Event Description
Information was received from a healthcare provider via a company representative regarding a patient receiving morphine 10mg/ml at 2mg/day and bupivacaine 40mg/ml at 8mg/day via an implantable pump.It was reported that the patient wasn¿t receiving medication through normal use of the pump, the patient wasn¿t getting medication as dosed.The catheter was checked for patency and found to not be patent.There were no environmental, external or patient factors that may have led or contributed to the issue.The diagnostics and troubleshooting performed was a dye study.A catheter revision was performed.It was found that the catheter connection was not secure resulting in ineffective delivery of medication.The pump was reconnected to catheter and found to be working as normal.The issue was resolved at the time of the report and it was noted that the healthcare provider would not have any further information regarding the event.
 
Manufacturer Narrative
Concomitant medical products: product id: 8780, serial#: (b)(4), implanted: (b)(6) 2022, product type: catheter.Other relevant device(s) are: product id: 8780, serial/lot #: (b)(4), ubd: 13-jan-2024, udi#: (b)(4).If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNCHROMED II
Type of Device
PUMP, INFUSION, IMPLANTED, PROGRAMMABLE
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
glen belmer
7000 central avenue ne rcw215
minneapolis, MN 55432
6122713209
MDR Report Key14495598
MDR Text Key292590472
Report Number3004209178-2022-06744
Device Sequence Number1
Product Code LKK
UDI-Device Identifier00643169630505
UDI-Public00643169630505
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P860004
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 05/25/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/25/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date10/14/2022
Device Model Number8637-20
Device Catalogue Number8637-20
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received05/21/2022
Date Device Manufactured04/17/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age54 YR
Patient SexMale
-
-