KARL LEIBINGER MEDIZINTECHNIK GMBH & CO. KG LEVEL ONE CMF; SCREW Back to Search Results
Model Number 25-878-05-91
Device Problems Adverse Event Without Identified Device or Use Problem (2993); Osseointegration Problem (3003)
Patient Problem Impaired Healing (2378)
Event Date 04/25/2022
Event Type  Injury  
Manufacturer Narrative
Multiple mdr reports were filed for this patient.Mdr: 9610905-2022-00023.
 
Event Description
It was reported that screws became loose due to patient grinding their teeth.They were removed and replaced.
 
Manufacturer Narrative
An investigation was performed in the lab and there were no indications of material or manufacturing defects.The complaint percentage was calculated, and it is determined that the complaint percentage falls within the design risk limits adhered to at klm.Review of the device history records was not possible due to no lot number being identified.The investigation results conclude that the root cause for this failure could not be determined.If further information is obtained that might add value to the contents of the investigation report, an additional follow-up report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEVEL ONE CMF
Type of Device
SCREW
Manufacturer (Section D)
KARL LEIBINGER MEDIZINTECHNIK GMBH & CO. KG
kolbinger strasse 10
muehlheim/donau, 78570
GM  78570
Manufacturer (Section G)
KLS MARTIN L.P.
p.o. box 16369
jacksonville FL 32245
Manufacturer Contact
jennifer damato
p.o. box 16369
jacksonville, FL 32245
9046417746
MDR Report Key14498335
MDR Text Key292802622
Report Number9610905-2022-00022
Device Sequence Number1
Product Code HRS
UDI-Device Identifier00888118047694
UDI-Public(01)00888118047694
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K971297
Exemption Number2017029
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 04/25/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/25/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model Number25-878-05-91
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/04/2022
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/16/2022
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age26 YR
Patient SexMale
Patient Weight115 KG
-
-