Catalog Number 1014261-080 |
Device Problems
Difficult to Advance (2920); Material Split, Cut or Torn (4008)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 05/12/2022 |
Event Type
malfunction
|
Event Description
|
It was reported that the procedure was to treat a femoral-popliteal by-pass with stenosis in the proximal part.A guide wire (gw) was mounted inside a 3x80mm armada 18 balloon catheter and advanced into the introducer sheath in the right groin.When the tip of the balloon catheter had reached the end of the introducer sheath, it was attempted to advance the gw through the balloon catheter to reach the target lesion but resistance was met when the gw had reached the tip of the balloon catheter.Via fluoroscopy, it was noted the gw had deviated from its central course inside the balloon catheter and was pointing to the side.Both the balloon catheter and gw were removed from the introducer sheath to inspect on the table.Once removed and inspected on the table, it was noted that the gw had deviated from its central course inside the balloon tip with the tip of the guidewire sticking out inside of the balloon.The procedure was completed using a drug-coated balloon catheter.There were no adverse patient effects and no clinically significant delay in the procedure.No additional information was provided.
|
|
Manufacturer Narrative
|
The device was not returned for analysis.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other similar incidents and/or complaints from this lot.The investigation was unable to determine a conclusive cause for the reported complaints.There is no indication of a product quality issue with respect to design, manufacture or labeling of the device.
|
|
Search Alerts/Recalls
|