Catalog Number 1014261-080
Device Problems Difficult to Advance (2920); Material Split, Cut or Torn (4008)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/12/2022
Event Type  malfunction  
Event Description
It was reported that the procedure was to treat a femoral-popliteal by-pass with stenosis in the proximal part.A guide wire (gw) was mounted inside a 3x80mm armada 18 balloon catheter and advanced into the introducer sheath in the right groin.When the tip of the balloon catheter had reached the end of the introducer sheath, it was attempted to advance the gw through the balloon catheter to reach the target lesion but resistance was met when the gw had reached the tip of the balloon catheter.Via fluoroscopy, it was noted the gw had deviated from its central course inside the balloon catheter and was pointing to the side.Both the balloon catheter and gw were removed from the introducer sheath to inspect on the table.Once removed and inspected on the table, it was noted that the gw had deviated from its central course inside the balloon tip with the tip of the guidewire sticking out inside of the balloon.The procedure was completed using a drug-coated balloon catheter.There were no adverse patient effects and no clinically significant delay in the procedure.No additional information was provided.
 
Manufacturer Narrative
The device was not returned for analysis.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other similar incidents and/or complaints from this lot.The investigation was unable to determine a conclusive cause for the reported complaints.There is no indication of a product quality issue with respect to design, manufacture or labeling of the device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARMADA 18 PTA CATHETER
Type of Device
PERIPHERAL DILATATION CATHETER
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 3005718570 (P099)
cashel road
clonmel tipperary
EI  
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key14503860
MDR Text Key292721232
Report Number2024168-2022-05628
Device Sequence Number1
Product Code LIT
Combination Product (y/n)N
Reporter Country CodeSW
PMA/PMN Number
K151317
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 05/25/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/26/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number1014261-080
Device Lot Number0120241
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received05/16/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/02/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-