• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON RESEARCH, LLC - ALCON PRECISION DEVICE MONARCH IOL DELIVERY SYSTEM, INJECTOR, UNSPECIFIED; LENS, GUIDE, INTRAOCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LLC - ALCON PRECISION DEVICE MONARCH IOL DELIVERY SYSTEM, INJECTOR, UNSPECIFIED; LENS, GUIDE, INTRAOCULAR Back to Search Results
Model Number ASKU
Device Problems Device Damaged by Another Device (2915); Failure to Eject (4010)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/02/2022
Event Type  malfunction  
Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Event Description
An other healthcare professional reported that, the plunger went over the lens when implanting and marked the iol.Lens was removed and back up used.Iol was explanted and back up implanted.Additional information received that the lens was not faulty, the injector plunger went over the lens when implanting.
 
Manufacturer Narrative
A sample was not received at the manufacturing site for evaluation for the report of plunger went over lens and marked iol; therefore, the condition of the product could not be verified.No lot number was identified with this complaint; therefore, a device history record review could not be conducted.A sample was not received at the manufacturing site and no lot information is available, therefore, the root cause for the customer complaint issue cannot be determined.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MONARCH IOL DELIVERY SYSTEM, INJECTOR, UNSPECIFIED
Type of Device
LENS, GUIDE, INTRAOCULAR
Manufacturer (Section D)
ALCON RESEARCH, LLC - ALCON PRECISION DEVICE
714 columbia avenue
sinking spring PA 19608
Manufacturer (Section G)
ALCON RESEARCH, LLC - ALCON PRECISION DEVICE
714 columbia avenue
sinking spring PA 19608
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key14515662
MDR Text Key292747254
Report Number2523835-2022-00190
Device Sequence Number1
Product Code KYB
Combination Product (y/n)N
Reporter Country CodeAS
PMA/PMN Number
UNK
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 09/02/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/26/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberASKU
Device Catalogue NumberASKU
Device Lot NumberASKU
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received08/07/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Treatment
ACRYSOF IQ VIVITY IOL; DUOVISC
-
-