• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION DREAMTOME RX 44; UNIT, ELECTROSURGICAL, ENDOSCOPIC (WITH OR WITHOUT ACCESSORIES)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION DREAMTOME RX 44; UNIT, ELECTROSURGICAL, ENDOSCOPIC (WITH OR WITHOUT ACCESSORIES) Back to Search Results
Model Number M00584040
Device Problems Break (1069); Failure to Fold (1255)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/01/2022
Event Type  malfunction  
Manufacturer Narrative
Date of event: approximated based on the date the manufacturer became aware of the event.The complainant was unable to provide the suspect device lot number; therefore, the lot expiration and device manufacture dates are unknown.(b)(4).The complainant indicated that the device was disposed and will not be returned for evaluation; therefore, a problem analysis of the complaint device could not be completed.If any further relevant information is identified, a supplemental medwatch will be filed.
 
Event Description
This report pertains to one of two dreamtome rx 44 used in the same procedure.It was reported to boston scientific corporation that a dreamtome rx 44 was used in the papilla during an endoscopic retrograde cholangiopancreatography (ercp) procedure.The exact procedure date was unknown.During the procedure, the cut wire "busted" when trying to make a cut.A second dreamtome rx 44 was used; however, the cut wire also "busted." it was reported that both the devices did not bow.The procedure was completed with another of the same device.There were no patient complications reported as a result of this event.No further information has been obtained despite good faith efforts.The reported event may suggest that the cutting wire broke.
 
Manufacturer Narrative
Block b3: approximated based on the date the manufacturer became aware of the event.Block d4, h4: the complainant was unable to provide the suspect device lot number; therefore, the lot expiration and device manufacture dates are unknown.Block h6: medical device problem code a0401 captures the reportable event of cutting wire broken.Conclusion code d17 is being used in lieu of an adequate conclusion code for device not returned.Block h10: the complainant indicated that the device was disposed and will not be returned for evaluation; therefore, a problem analysis of the complaint device could not be completed.If any further relevant information is identified, a supplemental medwatch will be filed.Block h2 (additional information): block b5 has been updated based on the additional information received on june 13, 2022 and on june 14, 2022.
 
Event Description
Note: this report pertains to one of two dreamtome rx 44 used in the same procedure.It was reported to boston scientific corporation that a dreamtome rx 44 was used in the papilla during an endoscopic retrograde cholangiopancreatography (ercp) procedure.The exact procedure date was unknown.During the procedure, the cut wire "busted" when trying to make a cut.A second dreamtome rx 44 was used; however, the cut wire also "busted." it was reported that both the devices did not bow.The procedure was completed with another of the same device.There were no patient complications reported as a result of this event.No further information has been obtained despite good faith efforts.The reported event may suggest that the cutting wire broke.Additional information received on june 13, 2022 and on june 14, 2022: it was reported to boston scientific corporation that a dreamtome rx 44 was used in the papilla during an endoscopic retrograde cholangiopancreatography (ercp) procedure.The exact procedure date was unknown.During the procedure, the cut wire "busted" when trying to make a cut.It was reported that the device did not bow and there was a break in the cutting wire.Additionally, no part of the cutting wire detached and fell into the patient.The procedure was completed with another of the same device.There were no patient complications reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DREAMTOME RX 44
Type of Device
UNIT, ELECTROSURGICAL, ENDOSCOPIC (WITH OR WITHOUT ACCESSORIES)
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
2546 calle primera
alajuela
CS  
Manufacturer Contact
carole morley
300 boston scientific way
marlborough, MA 01752
5086834015
MDR Report Key14516339
MDR Text Key293102683
Report Number3005099803-2022-02882
Device Sequence Number1
Product Code KNS
UDI-Device Identifier08714729778264
UDI-Public08714729778264
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K013153
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 06/28/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/26/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberM00584040
Device Catalogue Number8404
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received06/13/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-