• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OGDEN MANUFACTURING PLANT OPTIFLUX 180NRE DIALYZER FINISHED ASSY.; DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OGDEN MANUFACTURING PLANT OPTIFLUX 180NRE DIALYZER FINISHED ASSY.; DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM Back to Search Results
Model Number 0500318E
Device Problem Fluid/Blood Leak (1250)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/18/2022
Event Type  malfunction  
Event Description
A user facility administrator reported that a dialyzer blood leak occurred 30 minutes after the initiation of the patient¿s hemodialysis (hd) treatment.The blood leak was noted as being an internal blood leak.The leak was visually observed in the hansen lines.The machine, a fresenius 2008t bibag machine, alarmed appropriately with a blood leak alarm.Blood leak test strips were used and tested positive for the presence of blood.The patient¿s estimated blood loss (ebl) was approximately 300 ml.There was no patient injury, adverse events, or medical intervention required as a result of this event.The patient was restarted on a new machine and treatment completed successfully with new supplies.The complaint device was reported to not be available to be returned to the manufacturer for evaluation.
 
Manufacturer Narrative
The plant investigation is in process.A supplemental mdr will be submitted upon completion of this activity. .
 
Manufacturer Narrative
Plant investigation: the reported complaint was not confirmed as the complaint device was not returned for manufacturer evaluation.A production records review was performed on the reported lot.An investigation of the device history records (dhr) was conducted by the manufacturer.There was one approved temporary deviation notice (dn) reported on the lot which was unrelated to the complaint event.There was no indication of product nonacceptance, deviation, non-conformance, rework, labeling or process control failure during the manufacturing process which could be associated with the reported event.The lot met all release criteria.The reported lot number passed pyrogen testing, was within sterilization dosage parameters and passed all release criteria.There is no recall associated with this complaint.A definitive conclusion regarding the complaint incident cannot be reached without physical examination of the actual device.Therefore, the complaint is not confirmed.Continuous improvement is of the utmost importance to fresenius medical care as we strive to provide dialysis products of the highest quality to our patients.Reports of leaking product are investigated both individually as complaints, as well as via the nc/capa program, in order to assess and improve our products and processes.Capas for vision systems and blood leak reduction are recent examples of leak related investigations directed at an overall reduction in dialyzer leaks.
 
Event Description
A user facility administrator reported that a dialyzer blood leak occurred 30 minutes after the initiation of the patient¿s hemodialysis (hd) treatment.The blood leak was noted as being an internal blood leak.The leak was visually observed in the hansen lines.The machine, a fresenius 2008t bibag machine, alarmed appropriately with a blood leak alarm.Blood leak test strips were used and tested positive for the presence of blood.The patient¿s estimated blood loss (ebl) was approximately 300 ml.There was no patient injury, adverse events, or medical intervention required as a result of this event.The patient was restarted on a new machine and treatment completed successfully with new supplies.The complaint device was reported to not be available to be returned to the manufacturer for evaluation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OPTIFLUX 180NRE DIALYZER FINISHED ASSY.
Type of Device
DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM
Manufacturer (Section D)
OGDEN MANUFACTURING PLANT
director, site quality
475 west 13th street
ogden UT 84404
Manufacturer (Section G)
OGDEN MANUFACTURING PLANT
director, site quality
475 west 13th street
ogden UT 84404
Manufacturer Contact
jessica trujillo
920 winter st
waltham, MA 02451
6174175172
MDR Report Key14519399
MDR Text Key299459052
Report Number0001713747-2022-00179
Device Sequence Number1
Product Code KDI
UDI-Device Identifier00840861100156
UDI-Public00840861100156
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K162488
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Administrator/Supervisor
Type of Report Initial,Followup
Report Date 06/20/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/26/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number0500318E
Device Catalogue Number0500318E
Device Lot Number22CU01018
Was Device Available for Evaluation? No
Device AgeMO
Date Manufacturer Received06/16/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/30/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
FRESENIUS 2008T MACHINE; FRESENIUS 2008T MACHINE; FRESENIUS BLOODLINES; FRESENIUS BLOODLINES
Patient Age59 YR
Patient SexMale
Patient Weight64 KG
-
-