MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-712 PARADIGM INSULIN PUMP; PUMP, INFUSION, INSULIN
|
Back to Search Results |
|
Model Number MMT-712EWS |
Device Problems
Break (1069); Connection Problem (2900)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 05/20/2022 |
Event Type
malfunction
|
Event Description
|
Information received by medtronic indicated that the reservoir compartment cannot hold the reservoir.No harm requiring medical intervention was reported.The insulin pump will not be return for analyses.
|
|
Manufacturer Narrative
|
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.No conclusion can be drawn at this time.We therefore consider this report complete to the best of our knowledge.Medtronic, inc.(medtronic) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information in the time allotted and has provided as much information as is available to the company as of the submission date this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employees caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any "defects" or has "malfunctioned".These words are included in the fda 3500a form and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.Medtronic objects to the use of these words and others like it because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
|
|
Manufacturer Narrative
|
Pump received with multiple cracks on the case and broken reservoir tube lip.The pump passed the displacement test and the test infusion set, and reservoir does not lock into place due to the reservoir tube lip completely broken off.Broken reservoir tube lip, missing reservoir tube o ring, cracked case at the reservoir tube window corners, cracked reservoir tube window, cracked belt clip slot and broken battery tube threads.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employees caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
|
Search Alerts/Recalls
|
|
|