• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT VASCULAR TRAVELER RX CORONARY DILATATION CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT VASCULAR TRAVELER RX CORONARY DILATATION CATHETER Back to Search Results
Catalog Number 1013071-15
Device Problems Difficult to Remove (1528); Material Rupture (1546); Difficult to Advance (2920)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/09/2022
Event Type  malfunction  
Manufacturer Narrative
The device was not returned for evaluation. A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event. Additionally, a review of the complaint history of the reported lot did not indicate a lot specific quality issue. The investigation determined the reported complaints appear to be related to operational context. There is no indication of a product quality issue with respects to the design, manufacture, or labeling of the device. The traveler device is currently not commercially available in the us; however, it is similar to a device sold in the us.
 
Event Description
It was reported that the procedure was to treat a 90% stenosed distal right coronary artery (drca) with heavy calcification and moderate tortuosity. The 3. 50x15mm traveler balloon dilatation catheter (bdc) was advanced to the target lesion with difficulty due to the anatomy and the balloon ruptured at 6 atmospheres (atm) during the first inflation. The bdc was removed from the patient with difficulty also due to the anatomy. There was no adverse patient effect and no clinically significant delay reported. The procedure was completed with another traveler balloon. No additional information has been provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameTRAVELER RX CORONARY DILATATION CATHETER
Type of DeviceCORONARY DILATATION CATHETER
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR COSTA RICA, REG # 3009564766
52 calle 3 b31 coyol free zone
el coyol alajuela
CS
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key14569425
MDR Text Key293277796
Report Number2024168-2022-05860
Device Sequence Number1
Product Code LOX
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K103110
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative,Distributor
Reporter Occupation
Type of Report Initial
Report Date 06/01/2022
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received06/01/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Expiration Date10/31/2022
Device Catalogue Number1013071-15
Device Lot Number91120G1
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received05/09/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/20/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

-
-