Model Number 10742326
Device Problems Circuit Failure (1089); Electrical /Electronic Property Problem (1198)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/11/2022
Event Type  malfunction  
Event Description
It was reported to siemens that a malfunction occurred while operating the somatom force ct system.The user reported that the phs (patient table) did not move to the scanning start position after a topo scan (low dose overview scan).As a result, the male pediatric patient, who is less than one year in age, was scanned again on another ct system.It was reported that the patient is in good health and we are unaware of any impact to the state of health of the patient involved.
 
Manufacturer Narrative
Siemens has completed an investigation of the reported event.The root cause was determined to be a hardware error.A defective sp-wire-sensor has been identified and replaced.No general product issue could be identified.This report has been submitted with an abundance of caution.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SOMATOM FORCE
Type of Device
SYSTEM, X-RAY, TOMOGRAPHY, COMPUTED
Manufacturer (Section D)
SIEMENS HEALTHCARE GMBH-CT
siemensstr. 1
forchheim, germany 91301
GM  91301
Manufacturer (Section G)
SIEMENS HEALTHCARE GMBH-CT
siemensstr. 1
forchheim, germany 91301
GM   91301
Manufacturer Contact
meredith adams
40 liberty blvd.
malvern, PA 19355
4843231631
MDR Report Key14591568
MDR Text Key293265405
Report Number3004977335-2022-27639
Device Sequence Number1
Product Code JAK
UDI-Device Identifier04056869006956
UDI-Public04056869006956
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K190578
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility
Reporter Occupation Other
Remedial Action Repair
Type of Report Initial
Report Date 06/03/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/03/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number10742326
Was Device Available for Evaluation? Yes
Date Manufacturer Received05/11/2022
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age6 MO
Patient SexMale
-
-