Model Number URF-P7
Device Problem Material Integrity Problem (2978)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Event Description
An olympus employee reported a pinhole in the bending section cover of the subject device.The subject device was sent to an olympus service center for evaluation.During inspection and testing, service found the forceps stopper was scraped.This report is being submitted for the malfunction found during evaluation (damage to instrument channel).There was no harm or user injury reported due to the event.
 
Manufacturer Narrative
Inspection and testing found the forceps stopper was scraped due to physical stress.The reported event (pinhole) was confirmed during testing.There were leaks due to holes in the bending section cover caused by external factors.In addition, angulation was insufficient due to elongation of the angle wire, the insertion section was scratched due to external factors, and there were scratches on multiple parts of the device due to handling.The investigation is ongoing; therefore, the root cause of the reported event cannot be determined at this time.However, if additional information becomes available this report will be supplemented accordingly.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over 3 years since the device was manufactured.Based on the results of the investigation, it is likely that the reported phenomenon occurred due to stress of repeated use, external factors, or handling.A definitive root cause cannot be identified.Olympus will continue to monitor the field performance of this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
URETERO-RENO FIBERSCOPE
Type of Device
URETERO-RENO FIBERSCOPE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
Manufacturer Contact
kazutaka matsumoto
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8-507
JA   192-8507
426425177
MDR Report Key14595037
MDR Text Key301197494
Report Number8010047-2022-09389
Device Sequence Number1
Product Code FGB
UDI-Device Identifier04953170403811
UDI-Public04953170403811
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K181451
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 08/18/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/03/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberURF-P7
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/08/2022
Was the Report Sent to FDA? No
Date Manufacturer Received07/23/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/21/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-