WRIGHTS LANE SYNTHES USA PRODUCTS LLC DEPTH GAUGE FOR 2.0MM AND 2.4MM SCREWS; GAUGE, DEPTH
|
Back to Search Results |
|
Model Number 319.006 |
Device Problem
Break (1069)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 05/05/2022 |
Event Type
malfunction
|
Event Description
|
It was reported that on (b)(6) 2022, during inspection, the 2 broken small depth gauges and 2 broken screw forceps were found.There was no patient involvement.This report involves one (1) depth gauge for 2.0mm and 2.4mm screws.This is report 2 of 4 for (b)(4).
|
|
Manufacturer Narrative
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.The subject device has been received, the investigation is in progress, no conclusion could be drawn at the time of filing this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Manufacturer Narrative
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H10 additional narrative: d10 h3, h6: part: 319.006 synthes lot: 7582905 supplier lot: n/a release to warehouse date: 24 january 2014 manufactured by: synthes jennersville no nonconformance reports (ncrs) were generated during production.Review of the device history record(s) showed that there were no issues during the manufacture of this product, and any sub-components, which would contribute to this complaint condition.The product was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the returned device.Visual analysis of the returned sample revealed that the needle of the depth gauge for 2.0mm and 2.4mm screws has broken outside of the operating room, the broken needle was returned.No other problems was reported.A dimensional inspection was not performed for the depth gauge for 2.0mm and 2.4mm screws due to post manufacturing damage.The observed condition of the device was consistent with a random component failure that may have been caused by exposure to excessive/unintended forces.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was confirmed as the observed condition of the depth gauge for 2.0mm and 2.4mm screws would contribute to the complained device issue.There is no indication that a design or manufacturing issue has caused the complaint condition and hence the root cause cannot be determined.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Drawing/specifications reviewed current and manufactured revisions were reviewed.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.D9.
|
|
Search Alerts/Recalls
|
|
|