• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: WILLIAM COOK EUROPE COOK CELECT NAVALIGN JUGULAR & FEMORAL VENA CAVA FILTER SET; DTK FILTER, INTRAVASCULAR, CARDIOVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WILLIAM COOK EUROPE COOK CELECT NAVALIGN JUGULAR & FEMORAL VENA CAVA FILTER SET; DTK FILTER, INTRAVASCULAR, CARDIOVASCULAR Back to Search Results
Model Number G52915
Device Problems Obstruction of Flow (2423); Structural Problem (2506); Device Tipped Over (2589); Appropriate Term/Code Not Available (3191)
Patient Problems Internal Organ Perforation (1987); Pain (1994); Perforation of Vessels (2135); Stenosis (2263); Discomfort (2330); Insufficient Information (4580)
Event Type  Injury  
Event Description
It is alleged that the patient received a celect inferior vena cava (ivc) filter on (b)(6) 2014, and the patient was injured without further explanation.Hospital and medical records have been requested, but not yet provided.
 
Manufacturer Narrative
Blank fields on this form indicate the information is unknown or unavailable.Non-healthcare professional.Investigation: it has not been possible to further investigate or evaluate this alleged event based on the limited information and/or no device failure provided to date.Catalog number and lot number are unknown, however, the alleged celect is manufactured and inspected according to specifications.No evidence to suggest that this device was not manufactured according to specifications and nothing indicates that the filter did not perform as intended, e.G.Intended for the prevention of recurrent pulmonary embolism (pe) via placement in the vena cava.Cook will reopen its investigation if further information is received warranting supplementation in accordance with 21 c.F.R.803.56.This report includes information known at this time.A follow-up medwatch report will be submitted if additional relevant information becomes available.
 
Manufacturer Narrative
Investigation: the following allegations have been investigated: organ/vena cava/aorta perforation, stenosis, tilt, pain, discomfort.Investigation is reopened due to additional information provided.The reported allegations have been further investigated based on the information provided to date.Filter interacts with ivc wall, e.G.Penetration/perforation/embedment.This may be either symptomatic or asymptomatic.Potential causes may include improper deployment; and (or) excessive force or manipulations near an in-situ filter (e.G., a surgical or endovascular procedure in the vicinity of a filter).Potential adverse events that may occur include, but are not limited to, the following: trauma to adjacent structures, vascular trauma.Ivc occlusion/ thrombosis, new dvt, ivc stenosis as a reported complication, is a known risk in relation to filter implant and is well documented in the clinical literature and in clinical practice guidelines.This is supported by the clinical evidence report established to assess available clinical data to identify and evaluate the clinical safety and performance of the cook vena cava filters.Potential adverse events that may occur include, but are not limited to, the following: vena cava occlusion or thrombosis, vena cava stenosis, deep vein thrombosis.Filter tilt has been reported.Potential causes may include filter placement in ivcs with diameters larger than those specified in these instructions for use; improper deployment; manipulations near an implanted filter (e.G., a surgical or endovascular procedure in the vicinity of a filter); and (or) a failed retrieval attempt.Excessive filter tilt may contribute to difficult or failed retrieval; vena cava wall penetration/perforation; and (or) result in loss of filter efficiency.Potential adverse events that may occur include, but are not limited to, the following: unacceptable filter tilt.Unknown if the reported pain and discomfort is directly related to the filter and unable to identify a corresponding failure mode at this point in time.20 devices in lot.No relevant notes on work order.The product is manufactured and inspected according to specifications.No evidence to suggest that this device was not manufactured according to specifications, and nothing indicates that the filter did not perform as intended, e.G.Intended for the prevention of recurrent pulmonary embolism (pe) via placement in the vena cava.Cook will reopen its investigation if further information is received warranting supplementation in accordance with 21 c.F.R.803.56.This report includes information known at this time.A follow-up medwatch report will be submitted if additional relevant information becomes available.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, or that any cook device caused or contributed to or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Event Description
Patient allegedly received an implant on (b)(6) 2014 via the right internal jugular due to stroke.The patient alleges tilt, vena cava perforation, organ perforation, and stenosis.The patient further alleges pain and discomfort.(b)(6) 2020, per a report from computed tomography; ¿ivc filter with: tilting with the apex against the ivc wall.15 mm aortic, 11 mm small bowel, and 8 mm mesenteric perforation.Ivc stenosis.No fracture, or migration.¿.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COOK CELECT NAVALIGN JUGULAR & FEMORAL VENA CAVA FILTER SET
Type of Device
DTK FILTER, INTRAVASCULAR, CARDIOVASCULAR
Manufacturer (Section D)
WILLIAM COOK EUROPE
sandet 6
bjaeverskov 4632
DA  4632
Manufacturer (Section G)
WILLIAM COOK EUROPE
sandet 6, dk-4632
bjaeverskov
Manufacturer Contact
lissi walmann
sandet 6, dk-4632
bjaeverskov 
MDR Report Key14603667
MDR Text Key293367007
Report Number3002808486-2022-00770
Device Sequence Number1
Product Code DTK
UDI-Device Identifier10827002529158
UDI-Public(01)10827002529158(17)170507(10)E3212599
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Consumer
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 08/12/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/06/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/07/2017
Device Model NumberG52915
Device Catalogue NumberIGTCFS-65-1-UNI-CELECT
Device Lot NumberE3212599
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received07/21/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/01/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Life Threatening;
Patient SexMale
-
-