STRYKER TRAUMA KIEL PROXIMAL HUMERAL NAIL, CANNULATED, LEFT T2 PROX. HUM. Ø8 X 150 MM; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES Back to Search Results
Model Number 1832-1035S
Device Problem Material Twisted/Bent (2981)
Patient Problems Loss of Range of Motion (2032); Deformity/ Disfigurement (2360); Implant Pain (4561)
Event Date 02/08/2019
Event Type  Injury  
Manufacturer Narrative
This complaint has been generated based on findings discovered during post market surveillance literature review.The alleged pain, restricted range of motion (rom) due to failure of fixation with recurrent varus deformity, and slipping out of the proximal interlocking screw could not be confirmed, since the device was not returned for evaluation and no other additional information was received from the author.More detailed information about the patient medical history, the event circumstances, radiographs and the involved device(s) must be available in order to determine the root cause.If any additional information becomes available, the investigation will be reopened and re-evaluated accordingly.Device not available.
 
Event Description
The manufacturer became aware of a literature published by the ¿department of (b)(6) center, (b)(6) hospital, (b)(6)¿.The title of this report is, ¿evaluation of intramedullary nail fixation via the neviaser portal for proximal humerus fracture¿ published on february 08, 2019, which is associated with the stryker ¿t2 proximal humeral nailing system¿.The article can be found at https://doi.Org/10.1016/j.Jcot.2019.02.007.This report includes an analysis of the clinical data that was collected on 4 patients.The cases in this study range from april 2014 and march 2017.During the review of the literature, it was not possible to establish a specific device detail, patient information, and currently no additional device information is available.It was reported that 1 patient experienced pain, and restricted range of motion (rom) due to failure of fixation with recurrent varus deformity and slipping out of the proximal interlocking screw.The report states, ¿in this case, we used a bent nail unlike the other three cases in which straight nails were used.We consider this difference can be a cause of this complication.This patient also experienced failure of fixation with recurrent varus deformity and slipping out of the proximal interlocking screw at 6 weeks postoperatively.However, she remained under observation, because her pain was within the manageable limit, and finally, she was able to achieve bone-union without any exacerbation of her symptoms.Although active rom of her affected shoulder was restricted and moderate pain was still experienced during the shoulder motion, she was able to use her affected arm to perform daily living activities.¿.
 
Event Description
The manufacturer became aware of a literature published by the ¿department of orthopaedic hand and microsurgery center, osaka ekisaikai hospital, japan¿.The title of this report is, ¿evaluation of intramedullary nail fixation via the neviaser portal for proximal humerus fracture¿ published on february 08, 2019, which is associated with the stryker ¿t2 proximal humeral nailing system¿.The article can be found at https://doi.Org/10.1016/j.Jcot.2019.02.007.This report includes an analysis of the clinical data that was collected on 4 patients.The cases in this study range from april 2014 and march 2017.During the review of the literature, it was not possible to establish a specific device detail, patient information, and currently no additional device information is available.It was reported that 1 patient experienced pain, and restricted range of motion (rom) due to failure of fixation with recurrent varus deformity and slipping out of the proximal interlocking screw.The report states, ¿in this case, we used a bent nail unlike the other three cases in which straight nails were used.We consider this difference can be a cause of this complication.This patient also experienced failure of fixation with recurrent varus deformity and slipping out of the proximal interlocking screw at 6 weeks postoperatively.However, she remained under observation, because her pain was within the manageable limit, and finally, she was able to achieve bone-union without any exacerbation of her symptoms.Although active rom of her affected shoulder was restricted and moderate pain was still experienced during the shoulder motion, she was able to use her affected arm to perform daily living activities ¿.
 
Manufacturer Narrative
This complaint has been generated based on findings discovered during the post-market surveillance literature review.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.The alleged event could be confirmed, with the available radiographs in the literature.From the available radiographs, a formal medical opinion sought states: "that the approach used in surgery is not the one recommended in the op-tech for stryker t2 humeral nails.However, apart from this, such an indication can be treated with this nail.The nail in the x-ray seems to protrude beyond the cortex which may lead to a conflict with the acromion or the rotator cuff and result in reduced mobility.The issues may be related to the device, but it is more likely, that the complications that occurred are related to the procedure, from the x-ray it looks like the nail has not been positioned at the lateral center which would be necessary for a bent nail.The authors emphasize that it is possible, that this could be a general problem with the chosen entry point/portal/approach.Then it would be assessed as a procedure-related issue.With the available patient history, we learn, however, that there have been problems with moving already before the operation, making the insertion of the nail less likely to be the root cause of the described event.The proximal screws are backed out clearly.The nail stayed in place in the shaft part of the humerus, but it is possible, that the head dislocated slightly, when the fixation due to the backing out of the screws loosened.The dislocation and migration of the screws are far better visible than that of the nail".Based on the investigation and formal opinion, the root cause was attributed to off-label use.If any additional information becomes available, the investigation will be reopened and re-evaluated accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROXIMAL HUMERAL NAIL, CANNULATED, LEFT T2 PROX. HUM. Ø8 X 150 MM
Type of Device
ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES
Manufacturer (Section D)
STRYKER TRAUMA KIEL
prof. kuentscher-strasse 1-5
schoenkirchen/kiel D-242 32
GM  D-24232
Manufacturer (Section G)
STRYKER TRAUMA KIEL
prof. kuentscher-strasse 1-5
schoenkirchen/kiel D-242 32
GM   D-24232
Manufacturer Contact
anna jusinski
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key14610035
MDR Text Key293437868
Report Number0009610622-2022-00223
Device Sequence Number1
Product Code HSB
UDI-Device Identifier04546540515414
UDI-Public04546540515414
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K042396
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Literature
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 07/28/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/06/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Expiration Date12/31/2021
Device Model Number1832-1035S
Device Catalogue Number18321035S
Device Lot NumberK07C7BA
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received06/30/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/05/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
Patient SexFemale
-
-