DATEX-OHMEDA, INC. AISYS CS2; ANESTHESIA GAS MACHINE Back to Search Results
Device Problem Unexpected Shutdown (4019)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/10/2022
Event Type  malfunction  
Event Description
The hospital reported a system shutdown that resulted in a loss of mechanical ventilation during a case.There was no patient injury.
 
Manufacturer Narrative
Ge healthcare investigation into the reported occurrence is ongoing.A follow-up report will be issued when the investigation has been completed.Patient information could not be obtained due to country privacy laws.The initial reporter is located outside the u.S., and therefore this information is not provided due to country privacy laws.Unique identifier (b)(4).Legal manufacturer: (b)(4).Device evaluation anticipated, but not yet begun.
 
Manufacturer Narrative
A ge healthcare service representative performed a checkout of the system and confirmed the reported issue.The display cable was reseated to resolve the issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AISYS CS2
Type of Device
ANESTHESIA GAS MACHINE
Manufacturer (Section D)
DATEX-OHMEDA, INC.
3030 ohmeda dr,
madison, WI 53718
Manufacturer (Section G)
DATEX-OHMEDA, INC.
3030 ohmeda dr,
madison, WI 53718
Manufacturer Contact
anthony amenson
3030 ohmeda drive
madison, WI 53718
MDR Report Key14630841
MDR Text Key294133282
Report Number2112667-2022-01451
Device Sequence Number1
Product Code BSZ
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K170872
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 11/06/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/08/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received10/29/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/19/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage A
Patient Sequence Number1
-
-