Related manufacturer report number: 2017865-2022-11384, 2017865-2022-11385.During an in clinic follow-up, low sensing and high capture threshold were observed on the right ventricular (rv) and right atrial (ra) leads.The rv and ra leads were explanted and replaced to resolve the event.After the physician connected new leads, intermittent pacing was observed on the device.The patient was not pacemaker dependent and the device was explanted and replaced to resolve the event.The patient was in stable condition.
|
The reported event of intermittent pacing were not confirmed.Upon receipt, the device was in normal range of operation.The device image was analyzed and the average monthly voltage and current trends were normal.A visual analysis was performed on the header and no anomalies were observed.Additionally, telemetry, output and pacing function testing was performed and no anomalies were revealed.A longevity assessment was performed and device had appropriate longevity remaining.
|