• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITHS MEDICAL INTERNATIONAL, LTD. PORTEX ANESTHESIA BREATHING CIRCUITS; CIRCUIT, BREATHING (W CONNECTOR, ADAPTOR, Y PIECE)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITHS MEDICAL INTERNATIONAL, LTD. PORTEX ANESTHESIA BREATHING CIRCUITS; CIRCUIT, BREATHING (W CONNECTOR, ADAPTOR, Y PIECE) Back to Search Results
Catalog Number C1034J
Device Problem Gas/Air Leak (2946)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/01/2022
Event Type  malfunction  
Event Description
It was reported that during pre-use check of the device, leakage of air was observed from the product.The customer further checked it and found a crack in the corrugated tube.No additional information was available.No patient was involved.
 
Manufacturer Narrative
The device was returned, and the sample present a tier damage (hole) in the circuit.The circuit was connected and failed the leak test.The failure reported could not be reproduced using the tools from molding and assembly process.(updated h6) review of the device history record showed no issues related to the reported defect.The damage is believed to have occurred after the product left the manufacturing facility.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PORTEX ANESTHESIA BREATHING CIRCUITS
Type of Device
CIRCUIT, BREATHING (W CONNECTOR, ADAPTOR, Y PIECE)
Manufacturer (Section D)
SMITHS MEDICAL INTERNATIONAL, LTD.
boundary road
hythe, kent CT21 6JL
UK  CT21 6JL
Manufacturer (Section G)
SMITHS MEDICAL INTERNATIONAL, LTD.
boundary road
hythe, kent CT21 6JL
UK   CT21 6JL
Manufacturer Contact
jim vegel
6000 nathan lane north
7-1-1 akasaka,
minneapolis, MN 55442
MDR Report Key14700573
MDR Text Key295890539
Report Number3012307300-2022-11981
Device Sequence Number1
Product Code CAI
Combination Product (y/n)N
Reporter Country CodeJA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 09/08/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/15/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberC1034J
Device Lot Number4222992
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/23/2022
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received08/10/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/01/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-