During an atrial fibrillation procedure, an impedance issue occurred and the procedure was cancelled.When attempting to ablate the cavotricuspid isthmus cti without anatomy (just with the help of the cs catheter), 0w displayed on the generator several times and the impedance rose to 999.The ampere generator had to be restarted twice, grounded, and a second neutral electrode had to be used.A loose contact was also noticed on the flexability.The procedure was more complex than expected, and several maps and had to be carried out.The patient became restless over time.The system displayed several warnings, including "voxel collection is suspended because the system has detected an impedance shift.Existing voxels were deleted to resolve the issue.Then, the field frame disconnected and could not be reconnected.The neck patch also showed as disconnected.The physician did not want to validate the examination again or carry out further troubleshooting.The procedure had to be stopped because the anatomical position was not reached.There were no adverse patient consequences.It is suspected the issue was the patient's fault as no similar problems have occurred since.
|