Investigation of this event is pending and a supplemental report will be sent upon its completion.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
A supplemental report is being submitted for device evaluation and investigation completion.Product event summary: the ventricular assist device (vad) was returned for evaluation.No device malfunction was reported against (b)(6); therefore, the purpose of this investigation is solely to evaluate the device conformance to internal release requirements and/or to identify anomalies potentially introduced during use that may have prevented the pump from performing as intended.Review of the manufacturing records confirmed that the associated pump met all requirements for release.Log file analysis was not conducted since log files covering the reported event date were not available.Failure analysis of the returned pump revealed that the device passed visual examination and functional testing.Dimensional verification revealed that the front preload measurement, rear housing disc curvature, and front housing disc curvature were found to be deviating from release specifications.Capa pr00578223 was opened to investigate post-explant issues found during failure analysis of returned pumps.Internal pathological report revealed no evidence of thrombus within the device.Information received from the site indicated that the patient presented with sepsis, along with liver and renal failure.The patient had an external driveline infection and the vad was determined to be infected.Cultures confirmed the infection was fungal and bacterial, and antibiotics and intravenous fluids were administered.The patient's health continued to decline and the patient's family decided to transition the patient to comfort care; the vad was stopped and the patient died shortly after.Based on the available information, the device may have caused or contributed to the reported event.Per the instructions for use, infection, renal dysfunction, hepatic dysfunction, and death are known potential complications associated with the implantation of a vad.Based on review of past adverse events for this patient, it was noted that the patient had a history of infection and renal dysfunction.Possible clinical factors that may have contributed to this event include the patient¿s pre-existing history and related comorbidities, the progression of their underlying disease, and the patient's complex post-operative course, including care of the driveline exit site.There are possible patient, pharmacological and procedural factors that may have contributed to this event.Investigation of this event is completed and the file will be closed.If new information is received, the file will be re-opened and a supplemental will be submitted.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|