Model Number UM-S20-17S |
Device Problem
Leak/Splash (1354)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Event Description
|
The customer reported to olympus, the ultrasonic probe was flooded in mdr and did not work.It was reported that the device was not damaged during a procedure.There was no patient involvement, or user injury associated with this event.
|
|
Manufacturer Narrative
|
The customer suspected device was returned for investigation.The olympus service center confirmed the reported event during inspection and testing.Upon evaluation of the returned device, the following defects were found, leaking and damaged to ultrasonic insertion tube/probe and device was unable to produce ultrasonic image.The investigation is ongoing.A supplemental report will be submitted upon completion of the investigation or if any additional information is provided by the user facility.
|
|
Manufacturer Narrative
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.The investigation confirmed the reported device problem.The ultrasonic medium was leaking from a perforation/break in the sheath.As a result, the ultrasound image was not being produced.A review of the device history record found no deviations that could have caused or contributed to the reported device problem.A root cause for the leak/break in the sheath was not established.The occurrence of the reported problem can be prevented by adhering to the instructions for use (ifu), which state the following: ¿¿ never push or pull the ultrasonic probe with excessive force or withdraw it into the bending section of the endoscope during probe rotation (real-time mode).Manipulate the ultrasonic probe slowly and carefully when the endoscope is sharply angled or the forceps elevator is raised.Forcefully pushing or pulling the ultrasonic probe may damage it.¿ do not attempt probe rotation while the ultrasonic probe is advanced abruptly.This could result in image flow and impeded or cause irregular rotation.¿ olympus will continue to monitor field performance for this device.
|
|
Search Alerts/Recalls
|