• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SYNTHES GMBH TFNA FENESTRATED HELICAL BLADE 90MM - STERILE; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNTHES GMBH TFNA FENESTRATED HELICAL BLADE 90MM - STERILE; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES Back to Search Results
Catalog Number 04.038.390S
Device Problem Migration (4003)
Patient Problem Pain (1994)
Event Type  Injury  
Manufacturer Narrative
Additional narrative: patient id reported as (b)(6).Additional procode: ktt complainant part is not expected to be returned for manufacturer review/investigation.(b)(6).Without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
Device report from australia reports an event as follows: it was reported that patient was treated on (b)(6) 2020, for nof fracture with a trochanteric fixation nail advanced (tfna) 235mm nail, blade, and was augmented with traumacem.On (b)(6) 2021, the patient presented to clinic with progressive lateral thigh pain which had increased over the last twelve months.Upon imaging, the initial nof fracture appears to be healed but the tfna nail had fractured at the proximal aperture and the blade had migrated laterally.As of (b)(6) 2022, the patient still had lateral thigh pain but did not want further operations to remove the implants.This report is for a tfna fenestrated helical blade 90mm - sterile.This is report 2 of 2 for (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TFNA FENESTRATED HELICAL BLADE 90MM - STERILE
Type of Device
ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES
Manufacturer (Section D)
SYNTHES GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
Manufacturer (Section G)
SYNTHES SELZACH
bohackerweg 5
selzach 2545
SZ   2545
Manufacturer Contact
kate karberg
1302 wright lane east
west chester, PA 19380
3035526892
MDR Report Key14714720
MDR Text Key294936694
Report Number8030965-2022-04077
Device Sequence Number1
Product Code HSB
UDI-Device Identifier07611819652576
UDI-Public(01)07611819652576
Combination Product (y/n)N
Reporter Country CodeAS
PMA/PMN Number
K160167
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/16/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number04.038.390S
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received05/18/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Type of Device Usage Unknown
Patient Sequence Number1
Treatment
TFNA FEM NAIL Ø11 LE 125° L235 TIMO15; TRAUMACEM V+ CEMENT KIT 10 ML
Patient Outcome(s) Required Intervention;
Patient SexFemale
-
-