• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BLUE BELT TECHNOLOGIES RI BONE PINS 4 MM X 152 MM QTY: 2; ORTHOPEDIC STEREOTAXIC INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BLUE BELT TECHNOLOGIES RI BONE PINS 4 MM X 152 MM QTY: 2; ORTHOPEDIC STEREOTAXIC INSTRUMENT Back to Search Results
Model Number ROB10011
Device Problem Material Twisted/Bent (2981)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/01/2022
Event Type  malfunction  
Manufacturer Narrative
Internal complaint reference: (b)(4).
 
Event Description
It was reported that during a cori assisted tka surgery, four(4) ri ri bone pins 4 mm x 152 mm were bent and not entering smoothly into the bone, the surgeon had to force them.The procedure was resumed, without any delay, with s+n back-up devices.No patient complications were reported.
 
Manufacturer Narrative
H6.The ri bone pins 4 mm x 152 mm qty: (b)(4), part number rob10011, used for treatment was not returned for evaluation.A relationship between the reported event and the device could not be established.A complaint history review for similar reported/confirmed complaints has identified prior events.While all products meet required manufacturing specifications prior to release a serial number, lot number, or part revision is required to link the device to a dhr or nc investigation.The failure mode and associated risk have been anticipated within the risk file.The risk level is still adequate.Although the reported problem was not confirmed through a visual or functional evaluation, factors contributing to the reported symptom may have been associated with product mishandling during insertion or removal.Based on the investigation, the need for a corrective action is not recommended or required at this time.Should any additional information be received the complaint will be reopened.The failure mode will continue to be closely monitored through complaint investigation and trended through post market surveillance activities.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RI BONE PINS 4 MM X 152 MM QTY: 2
Type of Device
ORTHOPEDIC STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
BLUE BELT TECHNOLOGIES
2905 northwest blvd ste 40
plymouth MN 55441
Manufacturer (Section G)
BLUE BELT TECHNOLOGIES
2905 northwest blvd ste 40
plymouth MN 55441
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key14720997
MDR Text Key294642442
Report Number3010266064-2022-00500
Device Sequence Number1
Product Code OLO
UDI-Device Identifier00885556757307
UDI-Public00885556757307
Combination Product (y/n)N
Reporter Country CodeTC
PMA/PMN Number
K201022
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 08/22/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/16/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberROB10011
Device Catalogue NumberROB10011
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received08/22/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
REAL INTELLIGENCE CORI, SN:(B)(6)
-
-