• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. 630G INSULIN PUMP MMT-1715KL 630G; ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. 630G INSULIN PUMP MMT-1715KL 630G; ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND Back to Search Results
Model Number MMT-1715KL
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Hyperglycemia (1905)
Event Date 06/07/2022
Event Type  Injury  
Manufacturer Narrative
This mdr related to the puerto rico manufacturing site has been assigned a medwatch number from the medtronic minimed northridge site, per variance 5.Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.No conclusion can be drawn at this time.We therefore consider this report complete to the best of our knowledge.Medtronic, inc.(medtronic) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information in the time allotted and has provided as much information as is available to the company as of the submission date this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employees caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any "defects" or has "malfunctioned".These words are included in the fda 3500a form and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.Medtronic objects to the use of these words and others like it because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Event Description
The customer reported via phone call that they experienced high blood glucose.The customer¿s blood glucose level was 450 mg/dl.The customer¿s current blood glucose value was 270 mg/dl.The customer did not experience any symptoms of high blood glucose value.Auto mode feature was not activated at the time of high blood glucose event.Customer had been using the insulin pump system within 48 hours of reported high blood glucose event.No further complications were reported.The customer will continue to use the device.
 
Manufacturer Narrative
Retainer ring = black.Customer returned insulin pump for an alleged high blood glucoses found on (b)(6) 2022.The insulin pump passed the functional tests, including the self test, sleep current measurement, active current measurement, rewind test, prime/seating test, basic occlusion test, occlusion test, force sensor test, displacement test and delivery accuracy test at 0.08720 inches.The insulin pump was programmed with multiple bolus deliveries and all bolus delivered properly their indicated amounts (at quick bolus speed) and were properly recorded in the daily history.No bolus delivery anomaly or history anomaly noted.Successfully downloaded history files and traces using thus.¿ test p-cap and reservoir locked properly into reservoir compartment during testing.The following were noted during visual inspection: a pillowing keypad overlay and a scratched case.The insulin pump passed all the required testing.Unable to verify customer alleged for high blood glucoses.This mdr related to the puerto rico manufacturing site has been assigned a medwatch number from the medtronic minimed northridge site, per variance 5.Medtronic, inc.(medtronic) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information in the time allotted and has provided as much information as is available to the company as of the submission date this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employees caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any "defects" or has "malfunctioned".These words are included in the fda 3500a form and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.Medtronic objects to the use of these words and others like it because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
630G INSULIN PUMP MMT-1715KL 630G
Type of Device
ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
Manufacturer Contact
tricha miles
ceiba norte ind. park #50 road
juncos 00777--386
7635140379
MDR Report Key14725370
MDR Text Key294809101
Report Number2032227-2022-214569
Device Sequence Number1
Product Code OZO
UDI-Device Identifier000000763000316655
UDI-Public(01)000000763000316655(17)230515
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 09/23/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/17/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Expiration Date05/15/2023
Device Model NumberMMT-1715KL
Device Catalogue NumberMMT-1715KL
Device Lot NumberHG4BDFB
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Date Manufacturer Received09/22/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/15/2020
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Treatment
FRN-MMT-332-RSVR, UNOMED INF SET
Patient Outcome(s) Other;
Patient Age75 YR
Patient SexMale
Patient Weight104 KG
-
-