Catalog Number UNKNOWN- 2008T MACHINE
Device Problem Inadequate Ultra Filtration (1656)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/01/2022
Event Type  malfunction  
Event Description
A hemodialysis (hd) nurse reported a ultrafiltration (uf) issue that was occurring on unknown fresenius 2008t machines during hd treatments.Few details were obtained through follow up with a different nurse familiar with the recurring issue.It was unknown how many times it had occurred, and what machines were involved.During treatment, the uf goal on the machines was automatically resetting back to 3,000 ml and the uf time was resetting to 3 hours.It was unknown when the reset was occurring.Though specific details could not be provided, it was confirmed that no patients had any complaints or symptoms during or after their treatments.There were no adverse effects experienced, and no medical intervention was needed.The nurse did not know if any repairs were made to the machines.According to the nurse, the biomed does not come around often and works at multiple facilities.No additional information was provided.
 
Manufacturer Narrative
Plant investigation: the device was not returned to the manufacturer for physical evaluation and the serial number could not be obtained.Additionally, no on-site evaluation was performed by a fresenius field service technician (fst).As a serial number could not be determined, device history and manufacturing records could not be reviewed.As a physical evaluation could not be performed, a definitive conclusion regarding the reported incident could not be reached and a cause could not be confirmed.Should additional relevant information become available, a supplemental report will be submitted. .
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
2008T MACHINE
Type of Device
DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM
Manufacturer (Section D)
CONCORD MANUFACTURING
director, quality systems
4040 nelson avenue
concord CA 94520
Manufacturer (Section G)
CONCORD MANUFACTURING
director, quality systems
4040 nelson avenue
concord CA 94520
Manufacturer Contact
jessica trujillo
920 winter st
waltham, MA 02451
6174175172
MDR Report Key14735445
MDR Text Key302021999
Report Number0002937457-2022-00987
Device Sequence Number1
Product Code KDI
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K150708
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Nurse
Type of Report Initial
Report Date 06/17/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/17/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberUNKNOWN- 2008T MACHINE
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Device AgeMO
Date Manufacturer Received05/27/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-