• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CARDIOQUIP, LLC CARDIOQUIP MODULAR COOLER HEATER; CARDIOPULMONARY BYPASS DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CARDIOQUIP, LLC CARDIOQUIP MODULAR COOLER HEATER; CARDIOPULMONARY BYPASS DEVICE Back to Search Results
Model Number MCH-1000(I)
Device Problems Biofilm coating in Device (1062); Microbial Contamination of Device (2303)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/04/2019
Event Type  malfunction  
Manufacturer Narrative
The manufacturer is reporting the following complaint after a voluntary review of all complaints (reportable or not) since 2016.This report is being filed now, after being scrutinized under a newly revised risk matrix, recently adopted after inspection.The unit had discoloration in the tubing.The device met the criteria for bacterial contamination.The technician discovered multiple parts needing replacement as well as a cleaning procedure.During the investigation and servicing, we tested the water path surface and the results verified bacterial contamination in the unit as the cfu/ml count was greater than 738.0.After these parts were replaced and a cleaning and disinfection cycle was performed, the water path was tested.The results were 0.6 cfu/ml in the range of acceptable level, and functionality was restored.
 
Event Description
Customer is reporting discoloration in the tubing.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARDIOQUIP MODULAR COOLER HEATER
Type of Device
CARDIOPULMONARY BYPASS DEVICE
Manufacturer (Section D)
CARDIOQUIP, LLC
8422 calibration ct.
college station TX 77845
Manufacturer (Section G)
CARDIOQUIP, LLC
8422 calibration ct.
college station TX 77845
Manufacturer Contact
charley ford
8422 calibration ct.
college station, TX 77845
9796910202
MDR Report Key14739256
MDR Text Key301934783
Report Number3007899424-2022-00008
Device Sequence Number1
Product Code DWC
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K102147
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Biomedical Engineer
Remedial Action Repair
Type of Report Initial
Report Date 05/31/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/17/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberMCH-1000(I)
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/16/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-